Roche Diagnostics Corp. Roche VANC2, Online TDM Vancomycin, for use with the Roche/Hitachi Cobas c 501analyzer; Catalog No.04491050190. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche VANC2, Online TDM Vancomycin, for use with the Roche/Hitachi Cobas c 501analyzer; Catalog No.04491050190.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
Lots 144174, exp. 7/31/2007, 144195, exp. 9/30/2007 and 144207, exp. 11/30/2007.
Products Sold
Lots 144174, exp. 7/31/2007; 144195, exp. 9/30/2007 and 144207, exp. 11/30/2007.
Roche Diagnostics Corp. is recalling Roche VANC2, Online TDM Vancomycin, for use with the Roche/Hitachi Cobas c 501analyzer; Catalog No.0 due to Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
Recommended Action
Per FDA guidance
Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.