Roche Diagnostics Operations, Inc. Roche Elecsys Anti-CCP Immunoassay Catalog # 07251670190 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche Elecsys Anti-CCP Immunoassay Catalog # 07251670190
Brand
Roche Diagnostics Operations, Inc.
Lot Codes / Batch Numbers
exp 7/31/2020 467132, exp 9/30/2020
Products Sold
Lot # 376648 Exp : 08/31/2019 Lot # 388800 Exp : 09/30/2019 Anti-CCP assay: Anti-CCP Elecsys E2G 100 The affected lots of Anti-CCP have been distributed 02/28/2019 through 07/01/2019. ***Updated 7/17/20*** Current (in-date) and future lots of both Catalog numbers are impacted. In-date lots: 461913, exp 7/31/2020 467132, exp 9/30/2020
Roche Diagnostics Operations, Inc. is recalling Roche Elecsys Anti-CCP Immunoassay Catalog # 07251670190 due to Performance issues with certain lots of the Elecsys Anti-CCP assay on the cobas e 411 analyzer; MODULAR ANALYTICS E 170 module; and cobas e 601, 602, . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Performance issues with certain lots of the Elecsys Anti-CCP assay on the cobas e 411 analyzer; MODULAR ANALYTICS E 170 module; and cobas e 601, 602, and 801 modules with plasma samples.
Recommended Action
Per FDA guidance
Customers will be instructed to only use serum samples when testing with one of the affected lots numbers. Re-measure the sample if an implausible result occurs while using plasma samples, and a high Anti-CCP result does not match the patients clinical picture. To follow the work-around provided in the UMDC concerning pre-analytical sample handling for plasma samples. Perform maintenance according to the operator manual (e.g., Liquid Flow Cleaning [LFC]) to ensure proper functioning of the analyzer or module. Customers will be reminded that sample quality can be affected by fibrin clots and how this significantly impacts the Anti-CCP results. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC. ***Updated 7/17/20*** An updated letter was sent out to customers on July 13, 2020. After further investigation, Roche has decided to remove human plasma as a sample type from the intended use and remove all claims associated with the plasma sample type. The change is applicable for all current and upcoming lots. Customer actions required: " Use the Elecsys Anti-CCP assay with serum samples only for all current and future Elecsys Anti-CCP lot numbers. " ALL customers, complete the response form included with the letter and return to the firm " If you cannot run serum samples in your laboratory, complete the Product Information section of the response form to obtain reagent credit or contact your Roche sales representative for further options. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. " Discard Urgent Medical Device Correction TP-00714 if you previously received it.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, CA, FL, HI, IL, IN, IA, KY, LA, MA, MI, MO, MT, NJ, NY, OK, OR, PA, SC, TX, VA, PR