Roche Diagnostics Operations, Inc. Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers.
Brand
Roche Diagnostics Operations, Inc.
Lot Codes / Batch Numbers
CRPL3 - lot 42402301
Products Sold
CRPL3 - lot 42402301
Roche Diagnostics Operations, Inc. is recalling Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quanti due to Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c7. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
Recommended Action
Per FDA guidance
On November 6, 2020, the firm mailed customers Urgent Medical Device Correction letters. Customers were informed that the present communication serves as an update to previous UMDCs sent on 9/11/2019, 12/18/2019, and 8/27/2020, in that it includes the new product and lot number listed above. Customer Instructions: " Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. " Refer to the Product Replacement section on page 2 of the UMDC for product replacement information. " Complete all sections of the enclosed faxback form (TP-01128) and fax or email it according to the instructions on the form even if you are not requesting product replacement. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about this recall.