Roche Molecular Systems, Inc. cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190
Brand
Roche Molecular Systems, Inc.
Lot Codes / Batch Numbers
All lot numbers
Products Sold
All lot numbers
Roche Molecular Systems, Inc. is recalling cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190 due to False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate tre. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment
Recommended Action
Per FDA guidance
Roche issued Urgent Medical Device Correction letter on 9/3/21 via UPS Ground. Letter states reason for recall, health risk and action to take: Follow the cobas DNA Sample Preparation Kit (catalog number 05985536190) IFU for sample input with regard to the number of FFPET sections and DNA concentration. Follow the cobas cfDNA Sample Preparation Kit (catalog number 07247737190) IFU for plasma samples. " If an Ex20Ins Mutation Detected result is generated with the cobas EGFR Mutation Test v2, you must confirm the result with another method (e.g., sequencing or other PCR-based tests). " Consider reviewing results generated since May 2021. Amivantamab was approved by the U.S. Food and Drug Administration (FDA) for NSCLC patients with EGFR Ex20Ins on May 21, 2021. Test result reports with Ex20Ins Mutation Detected may be considered for confirmatory testing using sequencing or other PCR-based tests, upon the discretion of a CAP/CLIA laboratory director. " Consult with your facility physician and/or pathologist to determine any clinical implications specific to your patients. " Complete all sections of the enclosed faxback form (TP-01381) and fax it to 1-855-208-4038 or email it to Roche8166@sedgwick.com. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about the information contained in this UMDC
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.