SAMMONS PRESTON ROLYAN ValueLine Mat Platform Model C2230A,B,C; C922-30A, B,and C; C6904, C6903, C6902, C6901. All tables have the standard platform and the model numbers represent the additional shelves or drawers and the arrangement. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ValueLine Mat Platform Model C2230A,B,C; C922-30A, B,and C; C6904, C6903, C6902, C6901. All tables have the standard platform and the model numbers represent the additional shelves or drawers and the arrangement.
Brand
SAMMONS PRESTON ROLYAN
Lot Codes / Batch Numbers
Serial number prefixess 2509 throuth 2511.
Products Sold
Serial number prefixess 2509 throuth 2511.
SAMMONS PRESTON ROLYAN is recalling ValueLine Mat Platform Model C2230A,B,C; C922-30A, B,and C; C6904, C6903, C6902, C6901. All tables due to There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the ski. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
Recommended Action
Per FDA guidance
Letters were sent to customers certified mail on 1/25-26/2006 instructing them to take the tables out of service until it could be inspected by a sales representative. Sales represenatives are following up with customers in person to inspect tables.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026