Sarstedt Inc Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet)
Brand
Sarstedt Inc
Lot Codes / Batch Numbers
Product #85.1015.050, Lot #M40C219A7
Products Sold
Product #85.1015.050, Lot #M40C219A7
Sarstedt Inc is recalling Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet) due to The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated.
Recommended Action
Per FDA guidance
The consignee was notified via letter that was hand delivered on 08/27/2008. All of the distributed product was picked up from the consignee and returned to Sarstedt, Inc. None of the product had been used. All action was completed on 8/27/2008. Contact Sarstedt at 1-828-465-4000 for assistance.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
NC