SCC Soft Computer Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Dates: 4.0.2.38.1 06/15/07 4.02.39.2 06/29/07, 4.02.40.1 07/19/07, 4.02.41.1 08/20/07. The Result Reporting Interface is a highly versatile interface designed to electronically convey clinical laboratory orders and results for the SCC applications to other recipient information systems. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Dates: 4.0.2.38.1 06/15/07 4.02.39.2 06/29/07, 4.02.40.1 07/19/07, 4.02.41.1 08/20/07. The Result Reporting Interface is a highly versatile interface designed to electronically convey clinical laboratory orders and results for the SCC applications to other recipient information systems.
Brand
SCC Soft Computer
Lot Codes / Batch Numbers
Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1.
Products Sold
Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1.
SCC Soft Computer is recalling Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, due to SCC Soft Computer, Clearwater, FL issued a Correction for Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.0. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
SCC Soft Computer, Clearwater, FL issued a Correction for Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.0.2.40.1, and 4.0.2.41.1 on 10/05/2007. A software anomaly in the result reporting interface caused result comments entered after an empty line not to be sent to the HIS system. This included interpretative comments, results, and normal range c
Recommended Action
Per FDA guidance
SCC Soft Computer sent a RISK-TO-HEALTH NOTIFICATION TASK VERBIAGE dated October 5, 2007, through their Task Management System to all affected clients on October 17, 2007. The notice identified the product, the problem and the action clients should take to correct the problem. Clients were to acknowledge receipt of the task notice and to identify if the Workflow or functionality is not used and if they Agree to use alternative working solutions. Each client was asked to grant permission to load the required software to correct the issue. For any questions regarding this recall call (727) 789-0100.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, CO, FL, GA, IN, KS, KY, MA, MO, NH, NJ, NY, NC, OH, PA, TN, WA