SCC Soft Computer SoftID.NET versions 2.0.0.6 - 2.0.0.19. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftID.NET is to be used in a medical or clinical laboratory setting by knowledgeable, trained, and experienced professionals. SoftID.NET supports bedside collection of patient specimens used in diagnostic testing. This product is integrated with Soft Computer's SoftLab, SoftScape and can also interface with all other SCC laboratory systems and Hospital Information Systems (HIS). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SoftID.NET versions 2.0.0.6 - 2.0.0.19. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftID.NET is to be used in a medical or clinical laboratory setting by knowledgeable, trained, and experienced professionals. SoftID.NET supports bedside collection of patient specimens used in diagnostic testing. This product is integrated with Soft Computer's SoftLab, SoftScape and can also interface with all other SCC laboratory systems and Hospital Information Systems (HIS).
Brand
SCC Soft Computer
Lot Codes / Batch Numbers
Versions: SoftID.NET 2.0.0.6 - 4/3/2007, SoftID.NET 2.0.0.7 - 6/21/2007, SoftID.NET 2.0.0.8 - 8/29/2007, SoftID.NET 2.0.0.9 - 10/01/2007, SoftID.NET 2.0.0.10 - 10/25/2007, SoftID.NET 2.0.0.11 - 11/07/2007, SoftID.NET 2.0.0.12 - 2/08/2008, SoftID.NET 2.0.0.13 - 5/13/2008, SoftID.NET 2.0.0.14 - 6/12/2008, SoftID.NET 2.0.0.15 - 8/15/2008, SoftID.NET 2.0.0.16 - 10/15/2008, SoftID.NET 2.0.0.17 - 12/15/2008, SoftID.NET 2.0.0.18 - 2/02/2009, SoftID.NET 2.0.0.19 - 5/18/2009.
Products Sold
Versions: SoftID.NET 2.0.0.6 - 4/3/2007, SoftID.NET 2.0.0.7 - 6/21/2007, SoftID.NET 2.0.0.8 - 8/29/2007, SoftID.NET 2.0.0.9 - 10/01/2007, SoftID.NET 2.0.0.10 - 10/25/2007, SoftID.NET 2.0.0.11 - 11/07/2007, SoftID.NET 2.0.0.12 - 2/08/2008, SoftID.NET 2.0.0.13 - 5/13/2008, SoftID.NET 2.0.0.14 - 6/12/2008, SoftID.NET 2.0.0.15 - 8/15/2008, SoftID.NET 2.0.0.16 - 10/15/2008, SoftID.NET 2.0.0.17 - 12/15/2008, SoftID.NET 2.0.0.18 - 2/02/2009, SoftID.NET 2.0.0.19 - 5/18/2009.
SCC Soft Computer is recalling SoftID.NET versions 2.0.0.6 - 2.0.0.19. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 3376 due to SCC Soft Computer issued a Correction for SoftID.NET versions 2.0.0.6 - 2.0.0.19. When using the Cancel functionality, if the collection list refresh. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
SCC Soft Computer issued a Correction for SoftID.NET versions 2.0.0.6 - 2.0.0.19. When using the Cancel functionality, if the collection list refreshes during the transaction and changes the position of the patient on the list, the wrong patient order can be cancelled. All affected clients have been notified.
Recommended Action
Per FDA guidance
SCC Soft Computer notified Customers on August 28, 2009, via the proprietary communication software entitled "Risk-to-Health Notification Task Verbiage" dated August 17, 2009. The notification described the product, problem and action to be taken by customer. The customers were instructed to acknowledge receipt of the notification; document a request to load the software, if they wish to receive the correction; document that the Alternative Working Solution will be used by the customers facility until the correction has been taken live. If you have any questions, contact (727) 789-0100.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026