Sedecal USA, Inc Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain
Brand
Sedecal USA, Inc
Lot Codes / Batch Numbers
serial #PK10058
Products Sold
serial #PK10058
Sedecal USA, Inc is recalling Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro due to A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portab. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard.
Recommended Action
Per FDA guidance
An 'Important Safety Notice' dated 10/27/03 was sent to the one direct account on the same date, informing them of the non-compliances with the performance standard and requesting them to call Sedecal at 800-920-9525 to arrange for the modification of the x-ray unit at no cost to the user.