Shape Medical Systems, Inc Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions. Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions.
Brand
Shape Medical Systems, Inc
Lot Codes / Batch Numbers
Lot Numbers: 1412001, 1412002, 1412003 and 1412004
Products Sold
Lot Numbers: 1412001, 1412002, 1412003 and 1412004
Shape Medical Systems, Inc is recalling Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. due to This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis a. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The magnitude of error for VO2 or VCO2 parameters could be approximately 20% with use of the impacted product.
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, FL, GA, IN, MA, MN, NJ, NY, PA, TX, WI