Siemens Healthcare Diagnostics, Inc. Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
Brand
Siemens Healthcare Diagnostics, Inc.
Lot Codes / Batch Numbers
Lot Number: 211446 UDI: 00630414595566
Products Sold
Lot Number: 211446 UDI: 00630414595566
Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determin due to 2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Recommended Action
Per FDA guidance
Siemens issued An Urgent Medical Device Correction (UMDC) letter to US customers by FedEx beginning on 2021-12-01. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 2021-12-01. Letter states reason for recall, health risk and action to take: Discontinue use of and discard the kit lot listed in Table 1. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026