Atellica B2M Test (Siemens) – Result Accuracy Issue (2024)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
Brand
Siemens Healthcare Diagnostics, Inc.
Lot Codes / Batch Numbers
UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots
Products Sold
UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots
Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-m due to Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.
Recommended Action
Per FDA guidance
Siemens issued Urgent Medical Device Correction (UMDC) ACHC25-02.A.US was issued to US affected customers by FedEx beginning on 10/28/2024 and Urgent Field Safety Notice (UFSN) ACHC25-02.A.OUS. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Manually dilute patient samples that are > 18.00 mg/L: 1. Navigate to the CH Test Definition screen. 2. Under Measuring Interval, uncheck Repeat when Outside Measuring Interval. 3. For samples that generate test results > 18.00 mg/L, prepare a 1:2 manual dilution with saline solution (0.9% unbuffered Isotonic saline). 4. From the Create Patient Order screen, order the sample and enter 2 in the manual dilution field on the left of the screen. Note: The system will apply the 1:2 dilution factor and calculate the final result. 5. Process the manually diluted sample. "Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. "Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution The Atellica CH B2M assay IFU will be updated with the above steps to support manual 1:2 dilution only with saline for samples > 18.00 mg/L. Please see Appendix, Table 2, for data using these revised instructions. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.