Siemens Healthcare Diagnostics Inc. Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714
Brand
Siemens Healthcare Diagnostics Inc.
Lot Codes / Batch Numbers
Lot number GB9021 exp 2009-01-21 and Lot number GB9029 exp 2009-01-29
Products Sold
Lot number GB9021 exp 2009-01-21 and Lot number GB9029 exp 2009-01-29
Siemens Healthcare Diagnostics Inc. is recalling Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE due to False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and qu. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.
Recommended Action
Per FDA guidance
The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.