Siemens Healthcare Diagnostics, Inc. Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista System Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista System
Brand
Siemens Healthcare Diagnostics, Inc.
Lot Codes / Batch Numbers
Lot Number 9DM001
Products Sold
Lot Number 9DM001
Siemens Healthcare Diagnostics, Inc. is recalling Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM due to Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrati. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.
Recommended Action
Per FDA guidance
An Urgent Field Safety Notice dated August 31, 2009, was issued to all affected customers. Customers were instructed to discontinue the use of Dimension Vista System CHEM 1 Cal lot 9DM001 and discard any remaining product. If customers have any question they can contact Technical Solutions at 1-800-241-0420.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026