Siemens Healthcare Diagnostics Inc RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Brand
Siemens Healthcare Diagnostics Inc
Lot Codes / Batch Numbers
All units
Products Sold
All units
Siemens Healthcare Diagnostics Inc is recalling RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11 due to The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Recommended Action
Per FDA guidance
On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative