Siemens Healthcare Diagnostics Inc Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI).
Brand
Siemens Healthcare Diagnostics Inc
Lot Codes / Batch Numbers
Lot # 238330002 forward Expiration Dates - May 25, 2019 - November 2, 2019
Products Sold
Lot # 238330002 forward Expiration Dates - May 25, 2019 - November 2, 2019
Siemens Healthcare Diagnostics Inc is recalling Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the due to Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results.
Recommended Action
Per FDA guidance
On May 29, 2019, Siemens issued Urgent Medical Device Correction notices to U.S.A. customers via FedEx courier service and Urgent Field Safety Notices to international customers via email. Actions to be Taken by the Customer: 1. Do not use TestPaks within 2 mos of expiration. 2. Repeat samples with cTNL results above 0.07 ng/mL or your institution's established 99th percentile value. 3. Complete and return the Field Correction Effectiveness Check form 4. Retain a copy of the letter with your laboratory records and forward a copy of the notice to anyone to whom you may have distributed the product. Siemens is working to resolve this issue and will contact customers when additional information is available. Siemens will reimburse users for repeat tests and discarded cTNL TestPaks within 2 months of the printed expiration date associated with this field corrective action.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.