Siemens Medical Solutions USA, Inc (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira 10836777 & 10838610; MAGENTOM Free.Max 11408424 & 11408425; MAGNETOM Free.Star 11408426 & 11408427; MAGNETOM Mica 11292557; MAGNETOM Sempra 10840815 & 10840816; MAGNETOM Spectra 10655588 & 10837643; (2) Over Ear Headphones SMN: 11060845 used with the Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
(1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira 10836777 & 10838610; MAGENTOM Free.Max 11408424 & 11408425; MAGNETOM Free.Star 11408426 & 11408427; MAGNETOM Mica 11292557; MAGNETOM Sempra 10840815 & 10840816; MAGNETOM Spectra 10655588 & 10837643; (2) Over Ear Headphones SMN: 11060845 used with the
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
(1)Model/Material Number/UDI: MAGNETOM FREE.MAX 11408425 UDI: 04056869281742, MAGNETOM AMIRA 10836777 UDI: 04056869006819, MAGENTOM AMIRA 10838610 UDI: 04056869020488, MAGNETOM FREE.Max 11408424 UDI: 04056869281759, MAGNETOM Free.Star 11408426 UDI: 04056869281766, MAGNETOM Free.Star 11408427 UDI: 0405686981773, MAGNETOM Mica 11292557 NA
Products Sold
(1)Model/Material Number/UDI: MAGNETOM FREE.MAX 11408425 UDI: 04056869281742; MAGNETOM AMIRA 10836777 UDI: 04056869006819; MAGENTOM AMIRA 10838610 UDI: 04056869020488; MAGNETOM FREE.Max 11408424 UDI: 04056869281759; MAGNETOM Free.Star 11408426 UDI: 04056869281766; MAGNETOM Free.Star 11408427 UDI: 0405686981773; MAGNETOM Mica 11292557 NA; MAGNETOM Sempra 10840815 UDI: 04056869044521. Over Ear Headphones 11060845 (Serial Numbers 7981 to 32200Serial Numbers; 7981 to 32200 (2) Model UDI-DI MAGNETOM AERA 10432914 04056869006697 MAGNETOM ALTEA 11344915 04056869230719 11410371 04056869230726 MAGNETOM AVANTO FIT 11516216 04056869299266 MAGNETOM AVANTOFit 10849578 04056869006642 10849579 04056869006659 MAGENTOM CIMA.X 11647158 04056869975221¿ 11689304 04056869975245 MAGNETOM LUMINA 11344916 04056869230740 11516153 04056869260136 MAGENTOM PRISMA 10849582 04056869006727 MAGENTOM PRISMAFit 10849583 04056869006734 MAGNETOM SKYRA 10432915 04056869006703 MAGENTOM SKYRA 10849580 04056869006710 MAGENTOM SKYRA FIT 11516217 04056869299273 MAGNETOM SOLA 11291455 04056869164809 11410231 04056869217864 MAGNETOM SOLA FIT 11410482 04056869245195 MAGNETOM TERRA 10882764 04056869039190 MAGNETOM TERRA.X 11371477 04056869039190 MAGNETOM Viato.Mobile 11689652 04056869985558 MAGNETOM VIDA 11060815 04056869039176 11516152 04056869260143 MAGNETOM VIDAFit 11410481 04056869245188 Over Ear Headphones 11060845 (Serial Numbers 7981 to 32200
Siemens Medical Solutions USA, Inc is recalling (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Fr due to In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isoint. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used
Recommended Action
Per FDA guidance
Siemens Medical Solutions USA, Inc., issued Urgent Medical Device Correction Letter to affected US customers via MR078/23/S and MR072/23/S beginning on September 1, 2023. Customers with valid email addresses will be emailed a copy of the CSAN via Adobe GigaSign. Letter states reason for recall, health risk and action to take: Siemens Healthineers will correct the issue by exchanging the headband of the affected over-ear headphones at customer sites with the field update MR073/23/S. This replacement program will start in October 2023. With the exchange of the headband the root cause is eliminated. In parallel the manufacturing error has been corrected for any new over-ear headphones with serial numbers greater than 32200. These new headphones are not affected by the issue. If you have additional questions, please contact your local service organization, or call 1-800-888-7436. Acknowledge Receipt of this Customer Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification. If you have received this notification by email via Adobe GigaSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this customer notification by the US Postal Service or FedEx delivery, please send an email to recallsandrefusals.team@siemens-healthineers.com acknowledging that you have read and understand the content provided.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026