Siemens Medical Solutions USA, Inc 3D-I/III Ceiling Stand. Diagnostic x-ray tube mount Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
3D-I/III Ceiling Stand. Diagnostic x-ray tube mount
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Model numbers 0923/G0923, 1022, 1080, 1212687, 1371970, 1372325, 1654250, 1576164, 8186751, 7388002, 7389489, 1738509, 1738517, 1738525, 7276504, 7276512, 7276520, 7276553, 7465008, 7486822, and 8185530.
Products Sold
Model numbers 0923/G0923, 1022, 1080, 1212687, 1371970, 1372325, 1654250, 1576164, 8186751, 7388002, 7389489, 1738509, 1738517, 1738525, 7276504, 7276512, 7276520, 7276553, 7465008, 7486822, and 8185530.
Siemens Medical Solutions USA, Inc is recalling 3D-I/III Ceiling Stand. Diagnostic x-ray tube mount due to Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of i. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of injury to operators fingers.
Recommended Action
Per FDA guidance
The recalling firm has issued a field correction per Update Instructions AX012/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees to install a protective cover and additional warning labels on the device, beginning 04/28/2006.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026