Siemens Medical Solutions USA, Inc 3D TOP Ceiling Stand. X-Ray System model number 8773673 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
3D TOP Ceiling Stand. X-Ray System model number 8773673
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Model number 8773673
Products Sold
Model number 8773673
Siemens Medical Solutions USA, Inc is recalling 3D TOP Ceiling Stand. X-Ray System model number 8773673 due to screws on the rollers may become loose. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
screws on the rollers may become loose
Recommended Action
Per FDA guidance
The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026