Siemens Medical Solutions USA, Inc Artis pheno- Model No. 10849000 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Artis pheno- Model No. 10849000
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
UDI-DI: 04056869046877 Serial Numbers 164766 164140 164138 164287 164740 164210 164183 164184 164776 164312 164071 164034 164391 164763 164306 164735 164277 164095 164072 164771 164743 164048 164741 164241 164186 164069 164213 164728 164263 164060 164113 164398 164366 164279 164774 164074 164084 164086 164098 164153 164330 164389 164394 164772 164718 164163 164297 164778 164333 164377 164397 164300 164105 164227 164022 164250 164739 164703 164187 164211 164100 164097 164775 164303 164352 164726 164724 164760 164371 ***Updated 9/23/22*** 164794 164400 ***Updated 1/9/23*** 164793 164172 164795
Siemens Medical Solutions USA, Inc is recalling Artis pheno- Model No. 10849000 due to Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tub. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Recommended Action
Per FDA guidance
On August 10, 2022, the firm notified customers via "URGENT: MEDICAL DEVICE CORRECTION" Letter. Customers were advised that if the error occurs, the user should try to recover normal operation by a shutdown and restart of the system. Siemens will correct the error with a hardware modification via Update Instruction AX028/22/S and AX039/22/S. Siemens' service organization will contact customers shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. If you have any further questions, call 610-219-4834 or email: meredith.adams@siemens-healthineers.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.