Siemens Medical Solutions USA, Inc AXIOM Artis BA, Model Number 5904656 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
AXIOM Artis BA, Model Number 5904656
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial Numbers: 28159 28144 28154 28139 28155 28160 28156
Products Sold
Serial Numbers: 28159 28144 28154 28139 28155 28160 28156
Siemens Medical Solutions USA, Inc is recalling AXIOM Artis BA, Model Number 5904656 due to A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube wi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Recommended Action
Per FDA guidance
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.