Siemens Medical Solutions USA, Inc Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
S-N Affected systems: 10xxx: 10001, 10003, 10010 10027, 10029 10031, 10033 10040, 10042 10045, 10048 10052, 10054 10110, 10112 10119 11xxx: 11000 11004, 11006 11028, 11030 11101, 11103 11183, 11185 11205, 11207 11372, 11374 11387, 11600 11608 12xxx: 12000 12014, 12016 12024, 12026 12045, 12047 12052, 12054, 12200 12207, 12301 12493, 12495 12505, 12507 12522, 12524 12550, 12552 12571, 12573, 12575 12610, 12612 12746, 12800 12803, 12805 12811, 12813 12889, 12891 12959, 12961 12973, 12975 12999 13xxx: 13000 13085, 13087, 13088, 13092, 13103 13105, 13114, 13121 4xxxx: 416, 40000, 40001, 40003 40021, 40030, 40038, 40070 5xxxx: 50001 50053, 50055, 50075 Expanded Recall: 11143, 12490
Products Sold
S-N Affected systems: 10xxx: 10001; 10003; 10010 10027; 10029 10031; 10033 10040; 10042 10045; 10048 10052; 10054 10110; 10112 10119 11xxx: 11000 11004; 11006 11028; 11030 11101; 11103 11183; 11185 11205; 11207 11372; 11374 11387; 11600 11608 12xxx: 12000 12014; 12016 12024; 12026 12045; 12047 12052; 12054; 12200 12207; 12301 12493; 12495 12505; 12507 12522; 12524 12550; 12552 12571; 12573; 12575 12610; 12612 12746; 12800 12803; 12805 12811; 12813 12889; 12891 12959; 12961 12973; 12975 12999 13xxx: 13000 13085; 13087; 13088; 13092; 13103 13105; 13114; 13121 4xxxx: 416; 40000; 40001; 40003 40021; 40030; 40038; 40070 5xxxx: 50001 50053; 50055; 50075 Expanded Recall: 11143, 12490
Siemens Medical Solutions USA, Inc is recalling Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 1 due to Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator whi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Recommended Action
Per FDA guidance
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to affected customers via AX013/20/S dated July 14, 2020, The letter states reason for recall, health risk and action to take: We advise urgently to cease the use of the hot plugging feature until the pending hardware update AX014/20/S has been completed. Siemens will correct the hardware error via Update Instruction AX014/20/S in which a Siemens Service Engineer will replace the main control D80. Service organization will contact you shortly to arrange a date to performthis corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.