Siemens Medical Solutions USA, Inc. e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
Brand
Siemens Medical Solutions USA, Inc.
Lot Codes / Batch Numbers
all e.cam models: 4380213, 4380221, 5242826, 5977066, 5977074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7332625, 7332633, 7760809, 7760932, 7761161, 7823920, 7823938, 7823946, 7823953, 7823979, serial numbers 1 through 9091
Products Sold
all e.cam models: 4380213, 4380221, 5242826, 5977066, 5977074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7332625, 7332633, 7760809, 7760932, 7761161, 7823920, 7823938, 7823946, 7823953, 7823979; serial numbers 1 through 9091
Siemens Medical Solutions USA, Inc. is recalling e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; due to The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator.
Recommended Action
Per FDA guidance
Siemens sent Important Safety Customer Advisory letters dated 8/27/03 to all e.cam customers on 9/5-12/03. The accounts were informed of the problem with the detector''s radial brake not being applied properly, allowing slow, unintended downward motion of the detector that could result in injury to the patient or operator, and were advised not to leave the patient unattended while on the pallet and not to leave the patient under the detector upon completion of the study. The customers were advised that they would be contacted by their Siemens service provider to schedule a mandatory update to the e.cam to correct the problem. The upgrade will be done free of charge.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026