Siemens Medical Solutions USA, Inc LANTIS software with MedOnc Module; MedOnc module is an option of the LANTIS radiotherapy treatment planning software distributed as a component of the firm''s linear accelerator used to deliver radiation therapy to cancer patients. The MedOnc module is distributed by Siemens Medical Solutions USA Oncology Division, 4040 Nelson Avenue, Concord, CA 94520 with their linear accelerators software package called LANTIS. The MedOnc module option is distributed with all LANTIS software copies, but t Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
LANTIS software with MedOnc Module; MedOnc module is an option of the LANTIS radiotherapy treatment planning software distributed as a component of the firm''s linear accelerator used to deliver radiation therapy to cancer patients. The MedOnc module is distributed by Siemens Medical Solutions USA Oncology Division, 4040 Nelson Avenue, Concord, CA 94520 with their linear accelerators software package called LANTIS. The MedOnc module option is distributed with all LANTIS software copies, but t
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial #: 93-0001, 93-0016, 93-0027, 93-0289, 93-0304, 93-0306, 93-0333, 93-0368, 93-0430, 93-0436, 93-0549, 93-0639, 93-0641, 93-0652, 93-0664, 93-0703, 93-0718 and 93-0786
Products Sold
Serial #: 93-0001, 93-0016, 93-0027, 93-0289, 93-0304, 93-0306, 93-0333, 93-0368, 93-0430, 93-0436, 93-0549, 93-0639, 93-0641, 93-0652, 93-0664, 93-0703, 93-0718 and 93-0786
Siemens Medical Solutions USA, Inc is recalling LANTIS software with MedOnc Module; MedOnc module is an option of the LANTIS radiotherapy treatment due to When using the Creatine Clearance dose calculations a software bug will allow the algorithm to assume a patients age to be in years when it is entered. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When using the Creatine Clearance dose calculations a software bug will allow the algorithm to assume a patients age to be in years when it is entered in months. This miscalculation may result in patients less than two years old to be seriously overdosed with radiation.
Recommended Action
Per FDA guidance
The firm has sent out notification letters to the 18 sites with the activated MedOnc option. All 18 site had the letters delivered by Siemens Medical Solutions field service personnel by 1/6/2005.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026