Siemens Medical Solutions USA, Inc Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Serial Numbers: 82010 82038 82039 82040 82044 82054 82065 82066 82070 82102 82105 82108 82109 82205 82213 82217 82267 82308 82354 82355 82378 82383
Siemens Medical Solutions USA, Inc is recalling Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 due to Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).
Recommended Action
Per FDA guidance
Siemens issued Urgent Medical Device Correction notification letter dated 9/22/20 via AX053/20/S stating reason for recall, health risk and action to take: Siemens will correct the error with a hardware replacement of the Display Port transceiver via Update Instruction AX052/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.