Siemens Medical Solutions USA, Inc Siemens Cios Spin VA30-mobile X-Ray system, Material # 10308194 - Product Usage: The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siemens Cios Spin VA30-mobile X-Ray system, Material # 10308194 - Product Usage: The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Serial Numbers: 50124 50135 50131 50171 50129 50032 50033 50048 50050 50051 50073 50075 50081 50088 50098 50107 50109 50111 50120 50196 50102 50180 50136 50185 50096 50155 50091 50130 50127 50077 50078 50134 50122 50101 50092 50094 50166 50049 50192 50161 50125 50128 50132 50079 50160 50030
Siemens Medical Solutions USA, Inc is recalling Siemens Cios Spin VA30-mobile X-Ray system, Material # 10308194 - Product Usage: The Cios Alpha and due to Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the syst. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Recommended Action
Per FDA guidance
Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026