Siemens Medical Solutions USA, Inc Siemens Magnetom Sonata syngo, Model Number: 7106425, Siemens Medical Solutions USA, Malvern PA, 19355 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siemens Magnetom Sonata syngo, Model Number: 7106425, Siemens Medical Solutions USA, Malvern PA, 19355
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial Number: 21936
Products Sold
Serial Number: 21936
Siemens Medical Solutions USA, Inc is recalling Siemens Magnetom Sonata syngo, Model Number: 7106425, Siemens Medical Solutions USA, Malvern PA, 193 due to Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Recommended Action
Per FDA guidance
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.