Siemens Medical Solutions USA, Inc. Siemens Medical Solutions E.CAM Patient Handling System (PHS) Field of View (FOV) indicator bars; a component of the E.CAM Emission Computed Tomography System; part number 43 66 618 (right) and 43 81 088 (left). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siemens Medical Solutions E.CAM Patient Handling System (PHS) Field of View (FOV) indicator bars; a component of the E.CAM Emission Computed Tomography System; part number 43 66 618 (right) and 43 81 088 (left).
Brand
Siemens Medical Solutions USA, Inc.
Lot Codes / Batch Numbers
part number 43 66 618 (right) and 43 81 088 (left), serial numbers 00001 to 04031
Products Sold
part number 43 66 618 (right) and 43 81 088 (left), serial numbers 00001 to 04031
Siemens Medical Solutions USA, Inc. is recalling Siemens Medical Solutions E.CAM Patient Handling System (PHS) Field of View (FOV) indicator bars; a due to Patient fingers can get pinched if they attempt to pull themselves up from the patient handling system without any assistance and grabbing onto the fl. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Patient fingers can get pinched if they attempt to pull themselves up from the patient handling system without any assistance and grabbing onto the flip-up FOV indicator bars.
Recommended Action
Per FDA guidance
Siemens sent Urgent Field Correction Recall letters dated 12/18/08 to all users of the E.cam Systems. The accounts were informed of the potential risk to patients when they attempt to pull themselves up from the patient handling system (PHS) without any assistance and grabbing the field of view (FOV) indicator bar (aka flip-up armrests). The customers were advised to always assist the patient when getting on/off of the PHS. They also were advised that a Siemens Customer Service Representative would contact them within 90 days to schedule a corrective upgrade to the PHS with a modified FOV guide to mitigate the risk for patient injury free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance.