Siemens Medical Solutions USA, Inc Siemens Medical Solutions, ONCOR IMPRESSION Plus Digital Linear Accelerator, part number 5857912 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siemens Medical Solutions, ONCOR IMPRESSION Plus Digital Linear Accelerator, part number 5857912 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV.
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial Number: 3894
Products Sold
Serial Number: 3894
Siemens Medical Solutions USA, Inc is recalling Siemens Medical Solutions, ONCOR IMPRESSION Plus Digital Linear Accelerator, part number 5857912 equ due to Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Recommended Action
Per FDA guidance
Siemens sent a Customer Safety Advisory Notice dated 12/12/2007 to DEVA customers. The letter stated that Siemens would temporarily disable energies 5MeV and below for all customers that decide to continue using the DEVA, resulting in the temporary unavailability of energies 5MeV and below for all available. Customers wishing not to have these energies blocked must return their DEVA and they would be contacted by a local Siemens rep to schedule a service appointment. The letter stated that as a preventative measure that those customers should refrain from using the DEVA with any combination of energies 5MeV and below with field sizes smaller than 6cm x 6cm. The firm suggested using the fixed field electron applicators and cutouts with these energies and field sizes. Siemens will release a new control console software in 2008 for free-of-charge, that will limit the use of the DEVA to energies 6MeV and above, while allowing the use of lower energies using the fixed field electron applicatiors.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026