Siemens Medical Solutions USA, Inc Siregraph CF X-Ray System. Fluoroscopic image intensified. Model number 4466033 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Siregraph CF X-Ray System. Fluoroscopic image intensified. Model number 4466033
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial numbers: 1506, 1528, 1529, 1539, 1544, 1568, 1572, 1633, 1821, 1841, 1864, 1877, 2020, 2023, and 2126.
Products Sold
Serial numbers: 1506, 1528, 1529, 1539, 1544, 1568, 1572, 1633, 1821, 1841, 1864, 1877, 2020, 2023, and 2126.
Siemens Medical Solutions USA, Inc is recalling Siregraph CF X-Ray System. Fluoroscopic image intensified. Model number 4466033 due to X-ray tube movement when table is tilted more than 90 degrees. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
X-ray tube movement when table is tilted more than 90 degrees
Recommended Action
Per FDA guidance
The recalling firm issued a Customer DAfety Advisory letter dated 6/7/04 to the hospitals. The letter informs them of the potential problem and provides preventative measures to avoid its occurance.