Siemens Medical Solutions USA, Inc. Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Brand
Siemens Medical Solutions USA, Inc.
Lot Codes / Batch Numbers
part 8717741, serial numbers 1326, 1315, 1324, 1330, 1310, 1325, 1336, 1307, 1317, 1331, 1299, 1300, 1301, 1302, 1303, 1304, 1305, 1306, 1308, 1309, 1311, 1312, 1313, 1314, 1316, 1318, 1319, 1320, 1322, 1323, 1327, 1328, 1329, 1332, 1334, 1335, and 1337.
Products Sold
part 8717741, serial numbers 1326, 1315, 1324, 1330, 1310, 1325, 1336, 1307, 1317, 1331, 1299, 1300, 1301, 1302, 1303, 1304, 1305, 1306, 1308, 1309, 1311, 1312, 1313, 1314, 1316, 1318, 1319, 1320, 1322, 1323, 1327, 1328, 1329, 1332, 1334, 1335, and 1337.
Siemens Medical Solutions USA, Inc. is recalling Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions due to The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan.
Recommended Action
Per FDA guidance
Siemens sent Urgent Field Correction Recall letters dated 4/22/09 to the affected users of the Symbia S and Symbia T Systems. The accounts were informed of the potential faulty pinion gear which could crack and cause the gantry to spin freely, damaging the system and potentially injuring a patient during a scan. The customers were advised to stop using the system if they hear any abnormal grinding noise from the gantry. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a replacement of the pinion gear free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.