Siemens Medical Solutions USA, Inc syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected archive and V30A systems where image data was stored to LTS using software version older than V30A. Picture Archiving and communications system Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected archive and V30A systems where image data was stored to LTS using software version older than V30A. Picture Archiving and communications system
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Serial numbers: 1279, 1210, 1246, 1198, 1105, 1227, 1064, 1058, 1083, 1090, 1189, 1061, 1258, 1119, 1039, 1040, 1074, 1079, 1282 and 1056.
Products Sold
Serial numbers: 1279, 1210, 1246, 1198, 1105, 1227, 1064, 1058, 1083, 1090, 1189, 1061, 1258, 1119, 1039, 1040, 1074, 1079, 1282 and 1056.
Siemens Medical Solutions USA, Inc is recalling syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected arch due to An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configur. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configured to archive image data on a network file system (NFS) connected via a connected archive (Mount Point Solution).
Recommended Action
Per FDA guidance
A Customer Safety Advisory Notice, dated 02/09/2009 was sent to affected customers via Update Instruction IM013/09S. The letter informs customers of when the malfunction occurs and the potential issue, provides instructions to avoid its occurrence, and how the issue will be resolved. Customers are to ensure that the Safety Advisory is placed in the system's instructions for use and contact their SIEMENS Uptime Service Senter if they have any questions.