Siemens Medical Solutions USA, Inc Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression.
Brand
Siemens Medical Solutions USA, Inc
Lot Codes / Batch Numbers
Model number 7502029 - Serial numbers: 9620, 9621, 9622, 9623, 9624, 9625, 9630, 9631, 9632, 9633, 9634, 9639, 9640, 12068, 12271, 13090, 13358, 13428, 13480, 13637, 13727, 13742, 13945, 13971, 13979, 14059, 14086, and 14960.
Products Sold
Model number 7502029 - Serial numbers: 9620, 9621, 9622, 9623, 9624, 9625, 9630, 9631, 9632, 9633, 9634, 9639, 9640, 12068, 12271, 13090, 13358, 13428, 13480, 13637, 13727, 13742, 13945, 13971, 13979, 14059, 14086, and 14960.
Siemens Medical Solutions USA, Inc is recalling Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression. due to Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be nece. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or recovered from another source.
Recommended Action
Per FDA guidance
The recalling firm issued a Customer Safety Advisory notice dated 9/14/09 to all affected customers via Update Instructions IM100/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. To avoid the potential risk of the issue, the local compression function can be disabled as described in the letter. Depending on the image data, the disabling of compression will result in increased storage usage. The issue will be resolved in the release of version VA70B. For all potentially affected customers with "JPEG Compression" license an update instruction will be released which defines the necessary corrective actions in case of "Distorted" images. The update instruction will be performed on request by the customer service group. Further questions regarding this issue are directed to the local SIEMENS Uptime Service Center or +49 (9131) 84-2056.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, FL, GA, IL, IN, MD, MI, MO, NE, NY, NC, OH, PA, TN, TX, WV