Siemens Medical Solutions USA, Inc. V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options installed and is at software version VA10, VA10A, VA10B, VA10C, VA15, VA15A, VA15B, VA16 or VA16A, manufactured by Siemens Medical Solutions, Mountain View, CA. Multipurpose mobile software controlled diagnostic ultrasound system and on-screen display for thermal and mechanical indices related to bio-effect mechanisms. Its function is to acquire primary or secondary ultrasound echo data. The V5Ms is a m Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options installed and is at software version VA10, VA10A, VA10B, VA10C, VA15, VA15A, VA15B, VA16 or VA16A, manufactured by Siemens Medical Solutions, Mountain View, CA. Multipurpose mobile software controlled diagnostic ultrasound system and on-screen display for thermal and mechanical indices related to bio-effect mechanisms. Its function is to acquire primary or secondary ultrasound echo data. The V5Ms is a m
Brand
Siemens Medical Solutions USA, Inc.
Lot Codes / Batch Numbers
Model numbers: -10041461: Acuson S2000, -8263703: V5Ms Transducer, -10041524: S2000Cardiac Imaging and Quantification, -10436394: S2000 CV Imaging and Quantification, -10041525: S2000 Advanced Cardiac Analysis, -10041501: S2000 Cardiac Application Module, -10042966: S2000 Cardiac Application Module Upgrade, -10042990: S2000 Cardiac Application Module- No Physio Upgrade, -10042991: S2000 Card Application Module with Physio Upgrade, -100442992: S2000 Card App Module only Upgrade
Products Sold
Model numbers: -10041461: Acuson S2000; -8263703: V5Ms Transducer; -10041524: S2000Cardiac Imaging and Quantification; -10436394: S2000 CV Imaging and Quantification; -10041525: S2000 Advanced Cardiac Analysis; -10041501: S2000 Cardiac Application Module; -10042966: S2000 Cardiac Application Module Upgrade; -10042990: S2000 Cardiac Application Module- No Physio Upgrade; -10042991: S2000 Card Application Module with Physio Upgrade; -100442992: S2000 Card App Module only Upgrade
Siemens Medical Solutions USA, Inc. is recalling V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options instal due to If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used,. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal limit before automatic image freeze (and therefore power shut off) occurs at 43 degrees Centigrade.
Recommended Action
Per FDA guidance
Siemens sent a Customer Safety Advisory Notification letter on August 28, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their local Customer Service Engineer would install a software update as soon as it becomes available. Customers were instructed to pass the Customer Safety Advisory Notification letter to all who need to be aware of these issues until the corrective action is completed. For questions regarding this recall call 650-969-9112 or 800-422-8766.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.