Skytron, Div. The KMW Group, Inc Skytron General Purpose Surgical Table; Model 6600, Dist. by Skytron, Grand Rapids, MI. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Skytron General Purpose Surgical Table; Model 6600, Dist. by Skytron, Grand Rapids, MI.
Brand
Skytron, Div. The KMW Group, Inc
Lot Codes / Batch Numbers
All units. Serial numbers 6600-2D-001 through 6600-2L-102.
Products Sold
All units. Serial numbers 6600-2D-001 through 6600-2L-102.
Skytron, Div. The KMW Group, Inc is recalling Skytron General Purpose Surgical Table; Model 6600, Dist. by Skytron, Grand Rapids, MI. due to The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.
Recommended Action
Per FDA guidance
A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables.