Skytron, Div. The KMW Group, Inc VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Linx 300 Integration Control System Model: VGDF-SKY (version 2). For communication and visual image management in the operating room suite. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Linx 300 Integration Control System Model: VGDF-SKY (version 2). For communication and visual image management in the operating room suite.
Brand
Skytron, Div. The KMW Group, Inc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Serial Numbers: MXDM200501769 MXDM200501773 MXDM191100421 MXDM191100422 MXDM200200161 MXDM200200162 MXDM200200163 MXDM200200164 MXDM200501770 MXDM200501772 MXDM200501771
Skytron, Div. The KMW Group, Inc is recalling VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Li due to VGDF (VGA) Faceplate failure results in failure of video signal for SkyVision Linx 300, a secondary display. Video would continue on the primary sourc. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
VGDF (VGA) Faceplate failure results in failure of video signal for SkyVision Linx 300, a secondary display. Video would continue on the primary source device display
Recommended Action
Per FDA guidance
Skytron notified authorized representatives on September 11, 2020 via email states reason for recall and action to take: Please let us know if you have deployed these into a facility. We will provide new replacement VGA faceplates. Skytron is scheduled to receive the replacements from Opticis by Oct 12. Process: 1.)Please enter a no charge PO and you will receive an RMA for the faceplates. (returns@skytron.us) 2.)Skytron will send new faceplates. 3.)Complete & return the attached acknowledgement form. (This is an FDA recall and paperwork is essential!) 4.)Return original VGA faceplate to Skytron. The Customer Acknowledgement form states reason and health risk and will be completed for each facility signed by the Distributor Representative and Customer
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, MI, MS, NJ, PA