GS70H Salus Surgical Table (Skytron) – bluetooth control issue (2023)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GS70H Salus Surgical Table, Product Code 18-071-70
Brand
Skytron, LLC
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Serial Numbers:SUC-20207-0002 SUC-20208-0006 SUC-20209-0010 SUC-2020X-0014 SUC-2022Z-0018 SUC-20233-0022 SUC-20207-0003 SUC-20208-0007 SUC-20209-0011 SUC-2020X-0015 SUC-2022Z-0019 SUC-20233-0023 SUC-20208-0004 SUC-20209-0008 SUC-2020X-0012 SUC-2020X-0016 SUC-2022Z-0020 SUC-20233-0024 SUC-20208-0005 SUC-20209-0009 SUC-2020X-0013 SUC-2022Z-0017 SUC-2022Z-0021 SUC-20233-0025
Skytron, LLC is recalling GS70H Salus Surgical Table, Product Code 18-071-70 due to When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the contro. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.
Recommended Action
Per FDA guidance
An URGENT: Medical Device Correction dated 7/25/23 was sent to customers. Actions to be taken by the user: 1. Please promptly forward this notification if the device has been transferred to any other entity. 2. Please review appendices of affected serial and UDI numbers and identify devices within your facility. 3. Please do not use the table control in Bluetooth mode until the correction has been installed. 4. Please contact quality at quality@skytron.com with any questions. Skytron authorized service technicians will contact affected facilities to schedule this service. A location with a medical outlet and a power cord for the operating table available will be required to complete the corrective action. Skytron estimates that this correction activity will be completed by December 31, 2023. Please do not use the table control in Bluetooth mode until the correction has been installed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.