Smith & Nephew Inc Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116.
Brand
Smith & Nephew Inc
Lot Codes / Batch Numbers
Lot Nos.: 07EQT0003, 07EQT0004, 07EQT0005, 07JQT0003, 07KQT0002, 63207107, 07HM12429, and 07JM15582.
Products Sold
Lot Nos.: 07EQT0003, 07EQT0004, 07EQT0005, 07JQT0003, 07KQT0002, 63207107, 07HM12429, and 07JM15582.
Smith & Nephew Inc is recalling Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. M due to Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping.
Recommended Action
Per FDA guidance
Prior to the dissemination of the official notices, the firm notified personnel at our field offices of the BD syringe recall and advised that a subsequent recall of our Caption devices would follow. High volume users were contacted via phone on July 30th and the 31st and personnel at all affected field offices were notified via email on July 30, 2008 with an Urgent Medical Device Recall Letter. The firm requedsted that usere immediately cease distribution, examin the stock on hand and the stock of the accounts that have been suppplied and remove any affected ecices from all inventory. A verification notification was included with letter