Smith & Nephew, Inc. JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis
Brand
Smith & Nephew, Inc.
Lot Codes / Batch Numbers
UDI/DI 00885556176467, Batch Number 22BM17564
Products Sold
UDI/DI 00885556176467, Batch Number 22BM17564
Smith & Nephew, Inc. is recalling JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis due to The JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM.
Recommended Action
Per FDA guidance
Smith & Nephew, Inc. issued an Urgent Medical Device Recall Notice to its consignees on 09/27/2023 via email and overnight mail. The notice explain the issue, potential risk, and requested the following actions: Inspect inventory, locate, and quarantine the affected devices. Sales representatives, district offices, or distributors were directed to notify their customers The firm is seeking return of unused products.