Smith & Nephew, Inc. OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK-Absorbable Suture anchor Product Number: 72201992 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK-Absorbable Suture anchor Product Number: 72201992
Brand
Smith & Nephew, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Batch Numbers: 2012908 2016123 2017811 2018793 2018796 2019889 2023270 2024438 2025338 2027482 2029362 2032556 2033493 2034379 2034380 2034381 2036272 2036273 2038763 2050882 2050883 2049650 2049651 2053757 2056870
Smith & Nephew, Inc. is recalling OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK-Absorbable Suture anchor Product Number: 72201992 due to Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by th. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch
Recommended Action
Per FDA guidance
Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Verify this information on the official source