Smith & Nephew, Inc. Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 014771 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 014771
Brand
Smith & Nephew, Inc.
Lot Codes / Batch Numbers
Lot Number: 2042492 8/28/20 Expanded Recall- Lot number 2049408
Products Sold
Lot Number: 2042492 8/28/20 Expanded Recall- Lot number 2049408
Smith & Nephew, Inc. is recalling Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in due to The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due to the abse. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due to the absence of thesleeve
Recommended Action
Per FDA guidance
Smith & Nephew issued Urgent Medical Device Recall Notice R-2020-12 dated June 24, 2020 via email and FedEx. to distributors, sales reps, hospitals or surgeons. Letter states reason for recall, health risk and action to take: Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com. Expanded Recall: An updated FSN was sent on August 28, 2020 to customers that received the additional affected batch. The recall was expanded as a result of additional complaints.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026