Smith & Nephew, Inc. Tibial knee prosthesis Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Tibial knee prosthesis
Brand
Smith & Nephew, Inc.
Lot Codes / Batch Numbers
Model: 71420164, 71420166 and 71420184 Lot Numbers: 19MT31457, 19MT31458, 19MT31459, 20AT32611, 20AT32612, 20AT32613, 20AT33540, 20AT33541, 20AT33542, 20AT33543, 20AT33544, 20AT33545, 20AT33546, 20AT33547, 20AT33548, 20AT33569, 20AT33570, 20AT33571, 20BT33856, 20BT33857, 20BT33858, 20BT33859, 20BT33860, 20BT33861, 20BT33862, 20BT33863, 20BT33864, 20BT33865, 20BT33866, 20BT33878, 20BT33879, 20BT33880, 20BT33881, 20BT33882, 20BT33883, 20BT33884, 20BT33885, 20BT33886, 20BT33887, 20BT33888, 20BT33889, 20BT34357, 20BT34358, 20BT34359, 20BT34360, 20BT34379, 20BT34380, 20BT34381, 20BT34382, 20BT34383, 20BT34384, 20BT34385, 20BT34386, H2008427, G1923903, G1923904, G1923905, G1923933, G1923934, G1923935, G1923936, G1923964, G1923965, G1923966, G1923967, G1923992, G1923993, G1923994, G1924450, G1924451, G1924452, G1924453, G1924480, G1924481, G1924482, G1924483, G1924512, G1924513, G1924514, G1924543, G1924544, G1924545, G1924571, G1924572, G1924573, G1925072, G1925073, G1925074, G1925075, G1925102, G1925106, G1925107, G1925108, G1925109, G1925137, G1925138, G1925139, G1925140, G1925165, G1925166, G1925167, G1925168, G1925194, G1925195, G1925622, G1925623, G1925624, G1925625, G1925651, G1925652, G1925653, G1925654, G1925682, G1925683, G1925684, G1925685, G1925712, G1925713, G1925714, G1925715, G1925741, G1925742, G1926271, G1926272, G1926273, G1926274, G1926301, G1926302.
Products Sold
Model: 71420164; 71420166 and 71420184 Lot Numbers: 19MT31457, 19MT31458,19MT31459, 20AT32611, 20AT32612, 20AT32613, 20AT33540, 20AT33541, 20AT33542, 20AT33543, 20AT33544, 20AT33545, 20AT33546, 20AT33547, 20AT33548, 20AT33569, 20AT33570, 20AT33571, 20BT33856, 20BT33857, 20BT33858, 20BT33859, 20BT33860, 20BT33861, 20BT33862, 20BT33863, 20BT33864, 20BT33865, 20BT33866, 20BT33878, 20BT33879, 20BT33880, 20BT33881, 20BT33882, 20BT33883, 20BT33884, 20BT33885, 20BT33886, 20BT33887, 20BT33888, 20BT33889, 20BT34357, 20BT34358, 20BT34359, 20BT34360, 20BT34379, 20BT34380, 20BT34381, 20BT34382, 20BT34383, 20BT34384, 20BT34385, 20BT34386, H2008427, G1923903, G1923904, G1923905, G1923933, G1923934, G1923935, G1923936, G1923964, G1923965, G1923966, G1923967, G1923992, G1923993, G1923994, G1924450, G1924451, G1924452, G1924453, G1924480, G1924481, G1924482, G1924483, G1924512, G1924513, G1924514, G1924543, G1924544, G1924545, G1924571, G1924572, G1924573, G1925072, G1925073, G1925074, G1925075, G1925102, G1925106, G1925107, G1925108, G1925109, G1925137, G1925138, G1925139, G1925140, G1925165, G1925166, G1925167, G1925168, G1925194, G1925195, G1925622, G1925623, G1925624, G1925625, G1925651, G1925652, G1925653, G1925654, G1925682, G1925683, G1925684, G1925685, G1925712, G1925713, G1925714, G1925715, G1925741, G1925742, G1926271, G1926272, G1926273, G1926274, G1926301, G1926302.
Smith & Nephew, Inc. is recalling Tibial knee prosthesis due to Due to an inconsistency in the raw material process, specific lots may contain units with internal non-homogenous material defects.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to an inconsistency in the raw material process, specific lots may contain units with internal non-homogenous material defects.
Recommended Action
Per FDA guidance
On July 16th the firm send a letter to its consignees with the following instructions: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. Please provide a copy of the attached Physician Communication to the surgical staff. 3. If you have no product to return, please put an X in the appropriate location below. 4. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 5. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 6. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, AR, CT, FL, GA, IN, IA, KY, LA, MA, MI, MN, MO, NV, NJ, NY, NC, OH, PA, SC, TN, TX, WV, WI