Smiths Medical ASD Inc. CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-21 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-21
Brand
Smiths Medical ASD Inc.
Lot Codes / Batch Numbers
All serial numbers.
Products Sold
All serial numbers.
Smiths Medical ASD Inc. is recalling CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of due to Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retar. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
Recommended Action
Per FDA guidance
Smiths Medical issued an URGENT MEIDCAL DEVICE CORRECTION to its consignees on 07/16/2024 via traceable mailing. The notice explained the issue, potential risk and requested the following actions be taken: 1. Examine the external condition of the battery pack and look for evidence of damage to the outer case. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 2. Users with damaged battery packs should submit complaints to: globalcomplaints@icumed.com; 1-866-216-8806 3. Ensure all users or potential users of these products are immediately made aware of this notification. 4. Complete and return response form to smithsmedical5437@sedgwick.com. 5. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to smithsmedical5437@sedgwick.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026