Smiths Medical ASD Inc. NORMOFLO Irrigation Warming administration set (disposable), Models IR-40, IR-500, IR-600, IRI-600, IRI-600B, IR-700 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
NORMOFLO Irrigation Warming administration set (disposable), Models IR-40, IR-500, IR-600, IRI-600, IRI-600B, IR-700
Brand
Smiths Medical ASD Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Model IR-40: Product Code IR-40 Model IR-500: Product Code IR-500 Model IR-600: Product Code IR-600 Model IRI-600: Product Code IRI-600 Model IRI-600B: Product Code IRI-600B Model IR-700: Product Code IR-700
Smiths Medical ASD Inc. is recalling NORMOFLO Irrigation Warming administration set (disposable), Models IR-40, IR-500, IR-600, IRI-600, due to There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration o. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use.
Recommended Action
Per FDA guidance
Smiths Medical sent an Urgent Medical Device Recall Notice and attachments dated August 13, 2021 to all affected customers. The letter instructed all customers who purchased affected devices listed in the table on page 1 of the Urgent Medical Device Recall Notice must identify any of the affected products within their possession and refer to the detailed instructions on pages 2 and 3 of the notice. Distributors will be instructed to immediately forward a copy of the Urgent Medical Device Recall Notice and attachments to any customers to whom they have distributed affected product. Distributors will also be instructed to request that customers to who they have distributed affected product complete the Response Form and return it to the distributor. Distributors will indicate their identity as the distributor and document the consignees name and address. Questions regarding this recall notification may be forwarded to fieldactions@smiths-medical.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026