Smiths Medical MD, Inc. Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY.
Brand
Smiths Medical MD, Inc.
Lot Codes / Batch Numbers
Serial numbers affected: 691099, 691108, 691117, 691700, 691704, 691713, 691719, 691724, 691729, 691731, 691734, 691737, 691739, 691744, 691748, 691756, 691761, 691764, 691769, 691772, 691776, 691780, 691785, 691788, 691793, 691794, 691798.
Products Sold
Serial numbers affected: 691099, 691108, 691117, 691700, 691704, 691713, 691719, 691724, 691729, 691731, 691734, 691737, 691739, 691744, 691748, 691756, 691761, 691764, 691769, 691772, 691776, 691780, 691785, 691788, 691793, 691794, 691798.
Smiths Medical MD, Inc. is recalling Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory i due to The year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into an immediate 'Clock battery is Low/Service Immediately' or 'Clock Battery needs service soon' alarm message indicating that the pump requires servicing.
Recommended Action
Per FDA guidance
All consignees who have received the affected CADD-Prizm PCS II ambulatory infusion pumps and the CADD-Prizm VIP System ambulatory infusion pumps were contacted via telephone by February 1, 2007. The problem will be explained to the consignee and they will be asked to return the pumps to Smiths Medical MD in order for the problem to be corrected.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.