Smiths Medical MD, Inc. Deltec Cozmo Insulin Pump, Model 1800. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients but, can also be used in a healthcare facility. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Deltec Cozmo Insulin Pump, Model 1800. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients but, can also be used in a healthcare facility.
Brand
Smiths Medical MD, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Product Numbers: 21-1801-51 21-1801-81 21-1803-51 21-1803-81 21-1804-51 21-1804-81 21-1805-137 21-1805-138 21-1805-29 21-1805-49 21-1805-51 21-1805-81 21-1806-137 21-1806-138 21-1806-29 21-1806-49 21-1806-51 21-1806-81 21-1807-137 21-1807-138 21-1807-29 21-1807-49 21-1807-51 21-1807-81 21-1815-01 21-1815-135 21-1815-50 21-1816-01 21-1816-135 21-1816-50 21-1817-01 21-1817-135 21-1817-50 21-1825-81 21-1826-81 and 21-1827-81.
Smiths Medical MD, Inc. is recalling Deltec Cozmo Insulin Pump, Model 1800. Deltec Cozmo Insulin Pumps are ambulatory electromechanical due to Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 soft. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not accurately displayed by the Pump. The amount of extended Bolus displayed on the Pump Home Screen 2 and in the Bolus Summary Report is less than what w
Recommended Action
Per FDA guidance
Consignees were sent a Smiths Medical "Urgent Device Recall" letter dated November 20, 2008. The letter was addressed to Distributors, Clinicians, Patients and other Customers. The letter included a Description of the Problem, Advice on Action to be Taken by the User and requested the return of the Confirmation Form. In particular, users were given the option to check the "Replacement Pump" option on the Confirmation Form initiating the process for pump replacement with a recertified pump followed by a return of the defective product. The Confirmation Form can be faxed to 1-800-628-6322 in the U.S. or 1-651-628-7485 outside the U.S., completed at www.cozmore.com/notice or calling 1-800-501-5748 in the U.S.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY, DC