Smiths Medical PM, Inc. BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
Brand
Smiths Medical PM, Inc.
Lot Codes / Batch Numbers
3404002F/AI07010067/ AI07010068/
Products Sold
Catalog number/ serial number: 3404001/AI05110052/ AI05120000/ AI05120011/ AI05120012/ AI05120013/ AI05120018/ AI06060015/ AI06070047/ AI06100004/ AI06100005/ AI06100006/ , 3404001S/AI05110000/ AI05110001/ AI05110002/ AI05110003/ AI05110004/ AI05110005/ AI05110006/ AI05110007/ AI05110008/ AI05110009/ AI05110010/ AI05110011/ AI05110012/ AI05110013/ AI05110014/ AI05110015/ AI05110016/ AI05110017/ AI05110018/ AI05110019/ AI05110020/ AI05110021/ AI05110022/ AI05110023/ AI05110024/ AI05110025/ AI05110026/ AI05110027/ AI05110028/ AI05110029/ AI05110030/ AI05110031/ AI05110032/ AI05110033/ AI05110034/ AI05110035/ AI05110036/ AI05110037/ AI05110038/ AI05110039/ AI05120010/ AI05120016/ AI05120017/ AI05120019/ AI06010042/ AI06010043/ AI06060003/ AI06060004/ AI06060005/ AI06060006/ AI06060007/ AI06060008/ AI06060009/ AI06060010/ AI06060011/ , 3404002/AI05090021/ AI05090024/ AI05110049/ AI05120015/ AI06010035/ AI06010036/ AI06010049/ AI06040000/ AI06040001/ AI06040002/ AI06040003/ AI06040004/ AI06040005/ AI06040006/ AI06040007/ AI06040008/ AI06040009/ AI06040010/ AI06040011/ AI06040012/ AI06040013/ AI06040014/ AI06040015/ AI06040016/ AI06040017/ AI06040018/ AI06040020/ AI06060012/ AI06060013/ AI06060014/ AI06060017/ AI06060018/ AI06060019/ AI06060020/ AI06070037/ AI06070038/ AI06070039/ AI06070040/ AI06070042/ AI06070043/ AI06070044/ AI06070045/ AI06090053/ AI06100011/ AI06100012/ AI07010057/ AI07010058/ , 3404002E/AI05110040 AI05110041/ AI05110042/ AI05120001/ AI05120002/ AI05120005/ AI05120006/ AI05120007/ AI05120008/ AI05120009/ AI05120024/ AI06010041/ AI06010044/ AI06010045/ AI06010046/ AI06010048/ AI06040019/ AI06040023/ AI06040024/ AI06070041/ AI06070046/ AI06070048/ AI06070049/ AI06090050/ AI06090051/ , 3404002F/AI07010067/ AI07010068/ , 3404002G/AI05110050/ AI05110051/ AI05120003/ AI05120004/ AI06060021/ AI06060022/ AI06060023/ AI06070035/ AI06070036/ AI06090054/ AI06090055/ AI06090056/ AI06090057/ AI06090058/ AI06100008/ AI06100009/ AI06100010/ AI06100018/ , 3404002S/AI05110043/ AI05110044/ AI05110045/ AI06010037/ AI06010038/ AI06010039/ AI06010040/ AI06060000/ AI06060001/ AI06060002/ AI06060016/ AI06090052/ AI06100019/ AI06100020/ AI06100021/ AI06100022/ AI06100023/ AI06100024/ AI07010050/ AI07010051/
Smiths Medical PM, Inc. is recalling BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and due to Phantom pacer spikes caused by a damaged resistor.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Phantom pacer spikes caused by a damaged resistor.
Recommended Action
Per FDA guidance
Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
KY, LA, MS, PA, PR
Page updated: Jan 10, 2026