Sound Surgical Technologies, LLC Linemaster Wireless Footswitch Accessory, Model SP-9970214-005112010 and Model SP-9970214-005112232. Sound Surgical Technologies, LLC, Louisville, CO 80027. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Linemaster Wireless Footswitch Accessory, Model SP-9970214-005112010 and Model SP-9970214-005112232. Sound Surgical Technologies, LLC, Louisville, CO 80027.
Brand
Sound Surgical Technologies, LLC
Lot Codes / Batch Numbers
Model SP-9970214-005112010 and SP-9970214-005112232. All units.
Products Sold
Model SP-9970214-005112010 and SP-9970214-005112232. All units.
Sound Surgical Technologies, LLC is recalling Linemaster Wireless Footswitch Accessory, Model SP-9970214-005112010 and Model SP-9970214-005112232. due to Failure to shut-off: Surgical equipment, activated by a wireless foot switch with VASER system, may remain powered on when switch is no longer depres. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Failure to shut-off: Surgical equipment, activated by a wireless foot switch with VASER system, may remain powered on when switch is no longer depressed.
Recommended Action
Per FDA guidance
Consignees were sent an Advisory Notice on 12/28/2007. The notification describes the defect. Sound Surgical will be contacting facilities to arrange a service call to install new cables between the footswitch receiver and the VASER system. Facilities may contact the Technical Service Center at 1-888-677-8765.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, CA, CT, FL, GA, IN, KY, LA, MI, NJ, NY, OH, OK, PA, TX, VA