Southern Prosthetic Supply Co. SPS Blue Line Tube Clamp Adapters, Product No. BL120M Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SPS Blue Line Tube Clamp Adapters, Product No. BL120M
Brand
Southern Prosthetic Supply Co.
Lot Codes / Batch Numbers
Product number BL120M, with the letters, ''SPS'' etched into the product
Products Sold
Product number BL120M, with the letters, ''SPS'' etched into the product
Southern Prosthetic Supply Co. is recalling SPS Blue Line Tube Clamp Adapters, Product No. BL120M due to Device Fracturing: recent design and process changes leave the device susceptible to fracturing, which could cause the prosthesis to fail. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Device Fracturing: recent design and process changes leave the device susceptible to fracturing, which could cause the prosthesis to fail
Recommended Action
Per FDA guidance
On October 6, 2006, SPS faxed a recall notification to the 187 customers who had purchased one or more BL120M Tube Clamp Adapters from SPS between August 29 and September 20, 2006. On October 9, SPS contacted those customers that had purchased five or more Adapters via telephone to confirm they had received the faxed recall notification. On October 10, SPS mailed another copy of the recall notification, via certified mail, return receipt requested to all of its customers that had purchased one or more BL120M Tube Clamp Adapters. The customers were told to send the BL120M Adapter back to SPS following replacement.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026