Spacelabs Healthcare, Incorporated Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bags with an attached insert card Insert card reads "TruLink series Dual Temperature Cable REF 700-0032-00 Spacelabs Medical Issaquah, WA USA" Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bags with an attached insert card Insert card reads "TruLink series Dual Temperature Cable REF 700-0032-00 Spacelabs Medical Issaquah, WA USA"
Brand
Spacelabs Healthcare, Incorporated
Lot Codes / Batch Numbers
Lot numbers: OSC 01/07, OSC 02/07, OSC 03/07, OSC 04/07 and OSC 07/07.
Products Sold
Lot numbers: OSC 01/07, OSC 02/07, OSC 03/07, OSC 04/07 and OSC 07/07.
Spacelabs Healthcare, Incorporated is recalling Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bag due to Intermittent contact of temperature probe adaptors can cause a loss of temperature monitoring which can cause a delay in care when an alarm is missed.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Intermittent contact of temperature probe adaptors can cause a loss of temperature monitoring which can cause a delay in care when an alarm is missed.
Recommended Action
Per FDA guidance
Starting July 3, 2008 the firm issued a "MEDICAL DEVICE CORRECTION" letter notifying their customers of the potential hazards with the use of probe adapter, PN 700-031-00. All letters have a return receipt to verify each customer received the medical device correction letter. The firm will replace all PN 700-0031-00 adaptors with a replacement adaptor from their initial distributor. The firm initiated sending new adaptors separately, by mail, with a copy of the same Medical Device Correction letter to the firm's customers. The firm has requested their customers to throw away any adapter that is being recalled. Technical Support will be available for customer questions and technical questions and to ensure each unit is upgraded. Contact Spacelabs Healthcare at 425-657-7200 ext 5970 if you have questions.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026